Contrast-enhanced specific absorption rate-efficient 3D cardiac cine with respiratory-triggered radiofrequency gating

Henningsson M, Chan R, Goddu B, Goepfert L, Razavi R, Botnar R, Schaeffter T, Nezafat R. Contrast-enhanced specific absorption rate-efficient 3D cardiac cine with respiratory-triggered radiofrequency gating. J Magn Reson Imaging. 2013;37(4):986–92.

Abstract

PURPOSE: To investigate the use of radiofrequency (RF) gating in conjunction with a paramagnetic contrast agent to reduce the specific absorption rate (SAR) and increase the blood-myocardium contrast in balanced steady-state free precession (bSSFP) 3D cardiac cine. MATERIALS AND METHODS: RF gating was implemented by synchronizing the RF-excitation with an external respiratory sensor (bellows), which could additionally be used for respiratory gating. For reference, respiratory-gated 3D cine images were acquired without RF gating. Free-breathing 3D cine images were acquired in eight healthy subjects before and after contrast injection (Gd-BOPTA) and compared to breath-hold 2D cine. RESULTS: RF-gated 3D cine reduced the SAR by nearly 40% without introducing significant artifacts while providing left ventricle (LV) measurements similar to those obtained with 2D cine. The contrast-to-noise ratio (CNR) was significantly higher for 3D cine compared to 2D cine, both before and after contrast injection; however, no statistically significant CNR increase was observed for the postcontrast 3D cine compared to the precontrast acquisitions. CONCLUSION: Respiratory-triggered RF gating significantly reduces SAR in 3D cine acquisitions, which may enable a more widespread clinical use of 3D cine. Furthermore, CNR of 3D bSSFP cine is higher than of 2D and administration of Gd-BOPTA does not improve the CNR of 3D cine.
Last updated on 03/06/2023