Incidence, Treatment, and Outcomes of Symptomatic Device Lead-Related Venous Obstruction.

Ferro EG, Kramer DB, Li S, Locke AH, Misra S, Schmaier AA, Carroll BJ, Song Y, D’Avila AA, Yeh RW, Zimetbaum PJ, Secemsky EA. Incidence, Treatment, and Outcomes of Symptomatic Device Lead-Related Venous Obstruction. Journal of the American College of Cardiology. 2023;81(24):2328–2340. PMID: 37204378

BACKGROUND: The incidence and clinical impact of lead-related venous obstruction (LRVO) among patients with cardiovascular implantable electronic devices (CIEDs) is poorly defined.

OBJECTIVES: The objectives of this study were to determine the incidence of symptomatic LRVO after CIED implant; describe patterns in CIED extraction and revascularization; and quantify LRVO-related health care utilization based on each type of intervention.

METHODS: LRVO status was defined among Medicare beneficiaries after CIED implant from October 1, 2015, to December 31, 2020. Cumulative incidence functions of LRVO were estimated by Fine-Gray methods. LRVO predictors were identified using Cox regression. Incidence rates for LRVO-related health care visits were calculated with Poisson models.

RESULTS: Among 649,524 patients who underwent CIED implant, 28,214 developed LRVO, with 5.0% cumulative incidence at maximum follow-up of 5.2 years. Independent predictors of LRVO included CIEDs with >1 lead (HR: 1.09; 95% CI: 1.07-1.15), chronic kidney disease (HR: 1.17; 95% CI: 1.14-1.20), and malignancies (HR: 1.23; 95% CI: 1.20-1.27). Most patients with LRVO (85.2%) were managed conservatively. Among 4,186 (14.8%) patients undergoing intervention, 74.0% underwent CIED extraction and 26.0% percutaneous revascularization. Notably, 90% of the patients did not receive another CIED after extraction, with low use (2.2%) of leadless pacemakers. In adjusted models, extraction was associated with significant reductions in LRVO-related health care utilization (adjusted rate ratio: 0.58; 95% CI: 0.52-0.66) compared with conservative management.

CONCLUSIONS: In a large nationwide sample, the incidence of LRVO was substantial, affecting 1 of every 20 patients with CIEDs. Device extraction was the most common intervention and was associated with long-term reduction in recurrent health care utilization.

Last updated on 08/23/2023
PubMed