Publications

2024

Maher, Timothy R, Benjamin L Freedman, Shu Yang, Andrew H Locke, Robert D’Angelo, Madison Galvao, Alfred E Buxton, Jonathan W Waks, and Andre D’Avila. (2024) 2024. “Targeting Wavefront Discontinuity Lines for Scar-Related Ventricular Tachycardia Ablation: A Novel Functional Substrate Ablation Approach.”. JACC. Clinical Electrophysiology 10 (7 Pt 1): 1255-70. https://doi.org/10.1016/j.jacep.2024.03.023.

BACKGROUND: The boundaries of critical isthmuses for re-entrant ventricular tachycardia (VT) are formed by wavefront discontinuities (fixed lines of block, slow propagation, and rotational propagation) seen during baseline rhythm. It is unknown whether wavefront discontinuities can be automatically identified and targeted for ablation using electroanatomic mapping systems.

OBJECTIVES: The purpose of this study was to assess the electrophysiologic characteristics of automatically projected wavefront discontinuity lines (WADLs) and outcomes of an ablation strategy targeting WADLs in a mixed cohort of VT patients.

METHODS: Late activation substrate maps were analyzed from 1 or more baseline rhythm wavefronts. WADLs were identified using the Carto Extended Early Meets Late module. Number, total length, and distance to critical VT sites were measured. VT recurrence and VT-free survival were followed.

RESULTS: In total, 49 patients underwent 52 ablations with 71 unique substrate maps analyzed (18.8% epicardial; 62.0% right ventricular paced, 28.2% sinus rhythm, 9.9% left ventricular paced). A total of 28 VT critical sites were identified in 24 patients. WADLs were present in 49 of 71 (69.0%) maps. WADLs were present regardless of cardiomyopathy etiology, mapping wavefront, or surface. At a WADL threshold of 30%, 73.9% of critical VT sites were in close proximity (≤15 mm) to a WADL. VT-free survival was 62% at 1 year, with a competing risk model estimating a 1-year risk of VT recurrence of 23%.

CONCLUSIONS: WADLs can be automatically projected in a majority of patients in a mixed cohort of cardiomyopathy etiology, mapped wavefronts, and myocardial surfaces mapped. Targeting WADLs results in low rate of VT recurrence at 1 year.

2023

Aryana, Arash, Anna Sarcon, Mark R Bowers, Padraig Gearoid O’Neill, Maheer Gandhavadi, and Andre D’Avila. (2023) 2023. “Three-Dimensional Mapping, Recording and Ablation in Simulated and Induced Ventricular Tachyarrhythmias During Mechanical Circulatory Support Using the Percutaneous Heart Pump.”. Journal of Interventional Cardiac Electrophysiology : An International Journal of Arrhythmias and Pacing 66 (1): 99-107. https://doi.org/10.1007/s10840-021-01098-5.

PURPOSE: Due to their internal rotating magnets, conventional impeller-driven percutaneous ventricular assist devices (PVADs) yield high-frequency electrogram artifact and electromagnetic interference (EMI) when used with magnetic-based 3D electroanatomic mapping systems. The new percutaneous heart pump (PHP; Abbott, Chicago, IL) is a 14-French, 5-L/min, impeller axial-flow PVAD with a novel design that utilizes an external motor.

METHODS: We evaluated the feasibility of 3D mapping and radiofrequency ablation (RFA) in vivo during PHP mechanical circulatory support (MCS) in simulated ventricular tachycardia (pacing at 300 ms) and ventricular flutter (VFL, pacing at 200 ms) and also during ventricular fibrillation (VF) in a porcine model. Anterograde (right ventricular), transseptal, retrograde, and epicardial right and left ventricular 3D mapping (EnSite/CARTO) and RFA were performed in 6 swine using high-density mapping and force-sensing RFA catheters (TactiCath/ThermoCool). Surface and intracardiac electrograms and 3D maps were analyzed for noise/interference with and without MCS using PHP in sinus rhythm and simulated VT/VFL and VF.

RESULTS: Mapping and RFA proved feasible in the presence of MCS using PHP. The mean arterial pressure in sinus rhythm was 55 ± 2 mmHg (baseline) and 84 ± 4 mmHg during MCS with PHP and well-maintained during simulated VT (73 ± 8 mmHg) and VFL (65 ± 2 mmHg) and even in VF (65 ± 5 mmHg). No electrogram noise/artifact, EMI, or 3D map distortions were observed during mapping/RFA with either of two mapping systems.

CONCLUSIONS: Endocardial and epicardial 3D mapping and RFA in the presence of PHP are feasible and offer significant MCS during simulated VT/VFL and VF. Furthermore, PHP yielded no electrogram noise/artifact, EMI, or 3D mapping distortions in conjunction with magnetic-based 3D mapping systems.

Pérez, Juan J, Robert D’Angelo, Ana González-Suárez, Hiroshi Nakagawa, Enrique Berjano, and Andre D’Avila. (2023) 2023. “Low-Energy (360 J) Radiofrequency Catheter Ablation Using Moderate Power - short Duration: Proof of Concept Based on in Silico Modeling.”. Journal of Interventional Cardiac Electrophysiology : An International Journal of Arrhythmias and Pacing 66 (5): 1085-93. https://doi.org/10.1007/s10840-022-01292-z.

BACKGROUND: Pilot clinical studies suggest that very high power-very short duration (vHPvSD, 90 W/4 s, 360 J energy) is a feasible and safe technique for ablation of atrial fibrillation (AF), compared with standard applications using moderate power-moderate duration (30 W/30 s, 900 J energy). However, it is unclear whether alternate power and duration settings for the delivery of the same total energy would result in similar lesion formation. This study compares temperature dynamics and lesion size at different power-duration settings for the delivery of equivalent total energy (360 J).

METHODS: An in silico model of radiofrequency (RF) ablation was created using the Arrhenius function to estimate lesion size under different power-duration settings with energy balanced at 360 J: 30 W/12 s (MPSD), 50 W/7.2 s (HPSD), and 90 W/4 s (vHPvSD). Three catheter orientations were considered: parallel, 45°, and perpendicular.

RESULTS: In homogenous tissue, vHPvSD and HPSD produced similar size lesions independent of catheter orientation, both of which were slightly larger than MPSD (lesion size 0.1 mm deeper,   0.7 mm wider, and   25 mm3 larger volume). When considering heterogeneous tissue, these differences were smaller. Tissue reached higher absolute temperature with vHPvSD and HPSD (5-8 °C higher), which might increase risk of collateral tissue injury or steam pops.

CONCLUSION: Ablation for AF using MPSD or HPSD may be a feasible alternative to vHPvSD ablation given similar size lesions with similar total energy delivery (360 J). Lower absolute tissue temperature and slower heating may reduce risk of collateral tissue injury and steam pops associated with vHPvSD and longer applications using moderate power.

Irastorza, Ramiro M, Timothy Maher, Michael Barkagan, Rokas Liubasuskas, Enrique Berjano, and Andre D’Avila. (2023) 2023. “Anterior Vs. Posterior Position of Dispersive Patch During Radiofrequency Catheter Ablation: Insights from in Silico Modelling.”. Europace : European Pacing, Arrhythmias, and Cardiac Electrophysiology : Journal of the Working Groups on Cardiac Pacing, Arrhythmias, and Cardiac Cellular Electrophysiology of the European Society of Cardiology 25 (3): 1135-43. https://doi.org/10.1093/europace/euac277.

AIMS: To test the hypothesis that the dispersive patch (DP) location does not significantly affect the current distribution around the catheter tip during radiofrequency catheter ablation (RFCA) but may affect lesions size through differences in impedance due to factors far from the catheter tip.

METHODS: An in silico model of RFCA in the posterior left atrium and anterior right ventricle was created using anatomic measurements from patient thoracic computed tomography scans and tested the effect of anterior vs. posterior DP locations on baseline impedance, myocardial power delivery, radiofrequency current path, and predicted lesion size.

RESULTS: For posterior left atrium ablation, the baseline impedance, total current delivered, current distribution, and proportion of power delivered to the myocardium were all similar with both anterior and posterior DP locations, resulting in similar RFCA lesion sizes (< 0.2 mm difference). For anterior right ventricular (RV) ablation, an anterior DP location resulted in slightly higher proportion of power delivered to the myocardium and lower baseline impedance leading to slightly larger RFCA lesions (0.6 mm deeper and 0.8 mm wider).

CONCLUSIONS: An anterior vs. posterior DP location will not meaningfully affect RFCA for posterior left atrial ablation, and the slightly larger lesions predicted with anterior DP location for anterior RV ablation are of unclear clinical significance.

Aktaa, Suleman, Stylianos Tzeis, Chris P Gale, Michael J Ackerman, Elena Arbelo, Elijah R Behr, Lia Crotti, et al. (2023) 2023. “European Society of Cardiology Quality Indicators for the Management of Patients With Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death.”. Europace : European Pacing, Arrhythmias, and Cardiac Electrophysiology : Journal of the Working Groups on Cardiac Pacing, Arrhythmias, and Cardiac Cellular Electrophysiology of the European Society of Cardiology 25 (1): 199-210. https://doi.org/10.1093/europace/euac114.

To develop a suite of quality indicators (QIs) for the management of patients with ventricular arrhythmias (VA) and the prevention of sudden cardiac death (SCD). The Working Group comprised experts in heart rhythm management including Task Force members of the 2022 European Society of Cardiology (ESC) Clinical Practice Guidelines for the management of patients with VA and the prevention of SCD, members of the European Heart Rhythm Association, international experts, and a patient representative. We followed the ESC methodology for QI development, which involves (i) the identification of the key domains of care for the management of patients with VA and the prevention of SCD by constructing a conceptual framework of care, (ii) the development of candidate QIs by conducting a systematic review of the literature, (iii) the selection of the final set of QIs using a modified-Delphi method, and (iv) the evaluation of the feasibility of the developed QIs. We identified eight domains of care for the management of patients with VA and the prevention of SCD: (i) structural framework, (ii) screening and diagnosis, (iii) risk stratification, (iv) patient education and lifestyle modification, (v) pharmacological treatment, (vi) device therapy, (vii) catheter ablation, and (viii) outcomes, which included 17 main and 4 secondary QIs across these domains. Following a standardized methodology, we developed 21 QIs for the management of patients with VA and the prevention of SCD. The implementation of these QIs will improve the care and outcomes of patients with VA and contribute to the prevention of SCD.

Freedman, Benjamin L, Timothy R Maher, Madison Tracey, Pasquale Santangeli, and Andre D’Avila. (2023) 2023. “Procedural Adaptations to Avoid Haemodynamic Instability During Catheter Ablation of Scar-Related Ventricular Tachycardia.”. Arrhythmia & Electrophysiology Review 12: e20. https://doi.org/10.15420/aer.2022.24.

Classically, catheter ablation for scar-related ventricular tachycardia (VT) relied upon activation and entrainment mapping of induced VT. Advances in post-MI therapies have led to VTs that are faster and haemodynamically less stable, because of more heterogeneous myocardial fibrosis patterns. The PAINESD score is one means of identifying patients at highest risk for haemodynamic decompensation during attempted VT induction, who may, therefore, benefit from alternative ablation strategies. One strategy is to use temporary mechanical circulatory support, although this warrants formal assessment of cost-effectiveness. A second strategy is to minimise or avoid VT induction altogether by employing a family of 'substrate'-based approaches aimed at identifying VT isthmuses during sinus or paced rhythm. Substrate mapping techniques are diverse, and focus on the timing, morphology and amplitude of local ventricular electrograms - sometimes aided by advanced non-invasive cardiac imaging modalities. In this review, the evolution of VT ablation over time is discussed, with an emphasis on procedural adaptations to the challenge of haemodynamic instability.

Maher, Timothy R, Benjamin Freedman, Andrew H Locke, Madison Tracey, Jonathan W Waks, Diana Litmanovich, and Andre D’Avila. (2023) 2023. “Correlation Between Functional Substrate Mapping and Cardiac Computed Tomography-Derived Wall Thinning for Ventricular Tachycardia Ablation.”. JACC. Clinical Electrophysiology. https://doi.org/10.1016/j.jacep.2023.05.018.

BACKGROUND: Functional substrate mapping during baseline rhythm can identify arrhythmogenic tissue during ventricular tachycardia (VT) ablation. Wall thinning and wall thickness channels (WTCs) derived from computed tomography angiography have been shown to correlate with low voltage and VT isthmuses. The correlation between functional substrate mapping, wall thinning, and WTCs in patients with infarct- or non-infarct-related cardiomyopathies (ICM and NICM, respectively) has not been previously described.

OBJECTIVES: To correlate cardiac CTA-derived myocardial wall thinning with functional VT substrate mapping using isochronal late activation mapping.

METHODS: In 34 patients with ICM or NICM undergoing VT ablation who had a preprocedure computed tomography angiography, myocardial wall thinning was segmented in layers of 1 to 5 mm. Areas of wall thinning and WTCs were then spatially correlated with deceleration zones (DZs) from registered left ventricular endocardial isochronal late activation maps.

RESULTS: In 21 ICM patients and 13 NICM patients, ICM patients had greater surfaces areas of wall thinning (P < 0.001). In ICM patients, 94.1% of primary DZs were located on areas of wall thinning, compared to 20% of DZs in NICM patients overall but 50% if there was any wall thinning present. Fifty-nine percent of DZs in ICM patients and 56% of DZs in NICM patients were located near WTCs. The positive predictive value for WTC in localizing DZs was 22.5% and 37.8% in ICM and NICM patients, respectively.

CONCLUSIONS: Wall thinning is highly sensitive for functional substrate in ICM patients. WTCs had modest sensitivity for functional substrate but low positive predictive value for identifying DZs in ICM and NICM patients. These findings suggest that wall thinning may facilitate more efficient mapping in ICM patients, but WTCs are insufficient to localize wavefront discontinuities.

2022

Zimetbaum, Peter, Brett J Carroll, Andrew H Locke, Eric Secemsky, and Marc Schermerhorn. (2022) 2022. “Lead-Related Venous Obstruction in Patients With Implanted Cardiac Devices: JACC Review Topic of the Week.”. Journal of the American College of Cardiology 79 (3): 299-308. https://doi.org/10.1016/j.jacc.2021.11.017.

Cardiac implantable electronic device implantation rates have increased in recent decades. Venous obstruction of the subclavian, brachiocephalic, or superior vena cava veins represents an important complication of implanted leads. These forms of venous obstruction can result in significant symptoms as well as present a barrier to the implantation of additional device leads. The risk factors for the development of these complications remain poorly understood, and diagnosis relies on clinical recognition and cross-sectional imaging. Anticoagulation remains the mainstay of treatment, and thrombus debulking, lead extraction, venoplasty, and stenting are all important therapeutic interventions. This review provides a multidisciplinary-based approach to the evaluation and management of cardiac implantable electronic device lead-associated venous obstruction.

Irastorza, Ramiro M, Timothy Maher, Michael Barkagan, Rokas Liubasuskas, Juan J Pérez, Enrique Berjano, and Andre D’Avila. (2022) 2022. “Limitations of Baseline Impedance, Impedance Drop and Current for Radiofrequency Catheter Ablation Monitoring: Insights from In Silico Modeling.”. Journal of Cardiovascular Development and Disease 9 (10). https://doi.org/10.3390/jcdd9100336.

Background: Baseline impedance, radiofrequency current, and impedance drop during radiofrequency catheter ablation are thought to predict effective lesion formation. However, quantifying the contributions of local versus remote impedances provides insights into the limitations of indices using those parameters. Methods: An in silico model of left atrial radiofrequency catheter ablation was used based on human thoracic measurements and solved for (1) initial impedance (Z), (2) percentage of radiofrequency power delivered to the myocardium and blood (3) total radiofrequency current, (4) impedance drop during heating, and (5) lesion size after a 25 W−30 s ablation. Remote impedance was modeled by varying the mixing ratio between skeletal muscle and fat. Local impedance was modeled by varying insertion depth of the electrode (ID). Results: Increasing the remote impedance led to increased baseline impedance, lower system current delivery, and reduced lesion size. For ID = 0.5 mm, Z ranged from 115 to 132 Ω when fat percentage varied from 20 to 80%, resulting in a decrease in the RF current from 472 to 347 mA and a slight decrease in lesion size from 5.6 to 5.1 mm in depth, and from 9.2 to 8.0 mm in maximum width. In contrast, increasing the local impedance led to lower system current but larger lesions. For a 50% fat−muscle mixture, Z ranged from 118 to 138 Ω when ID varied from 0.3 to 1.9 mm, resulting in a decrease in the RF current from 463 to 443 mA and an increase in lesion size, from 5.2 up to 7.5 mm in depth, and from 8.4 up to 11.6 mm in maximum width. In cases of nearly identical Z but different contributions of local and remote impedance, markedly different lesions sizes were observed despite only small differences in RF current. Impedance drop better predicted lesion size (R2 > 0.93) than RF current (R2 < 0.1). Conclusions: Identical baseline impedances and observed RF currents can lead to markedly different lesion sizes with different relative contributions of local and remote impedances to the electrical circuit. These results provide mechanistic insights into the advantage of measuring local impedance and identifies potential limitations of indices incorporating baseline impedance or current to predict lesion quality.