TECTONIC CAD: Pivotal IDE Evaluation of Coronary Intravascular Lithotripsy for Severely Calcified Lesions.

Secemsky EA, Ali ZA, Shlofmitz E, Chehab BM, Jeremias A, Nicholson W, Sandesara P, Sayfo S, Verreault-Julien L, Latib A, Haddad C, Haddad R, Stoler RC, Osborn EA, Abela O, Smith J, Manthripragada G, Ben-Dor I, Hashim H, Kiss D, Rosenberg J, Syed M, Sandoval Y, Beohar N, Allana SS, Mannino S, Schmidt T, Mehrle A, Flaherty JD, Mukarram O, Choi C, Dohad S, Moussa I, Chester R, Popma JJ, Maehara A, Johnson G, Kraemer C, Pingle SC, Martinsen BJ, Meinen JM, Korngold EC, Kearney KE. TECTONIC CAD: Pivotal IDE Evaluation of Coronary Intravascular Lithotripsy for Severely Calcified Lesions.. JACC. Cardiovascular interventions. 2026; PMID: 42969900

Abstract

BACKGROUND: Severe coronary calcification complicates percutaneous coronary intervention by limiting lesion compliance and impairing stent expansion.

OBJECTIVES: To evaluate the safety and effectiveness of the SonicForce Coronary Intravascular Lithotripsy System (Abbott Vascular) in severely calcified de novo coronary lesions.

METHODS: TECTONIC CAD intravascular lithotripsy investigational device exemption (NCT06885177) was a prospective, multicenter, single-arm study of 335 patients at 40 U.S.

CENTERS: The primary safety endpoint was 30-day target lesion failure, defined as cardiac death, target-vessel myocardial infarction, or clinically driven target lesion revascularization. The primary effectiveness endpoint was procedural success. Each endpoint was tested against a prespecified performance goal. A prespecified optical coherence tomography (OCT) substudy evaluated calcium modification and stent expansion.

RESULTS: A total of 335 patients were included in the full analysis set, with 330 and 334 as the evaluable populations for the primary safety and primary effectiveness endpoints, respectively, and a median follow-up duration of 36 days (Q1-Q3: 28-37 days). Thirty-day target lesion failure was 0.3% (1 of 330; exact 2-sided 95% CI: 0.0%-1.7%); the upper 1-sided 97.5% confidence bound was 1.7%, below the 12.9% performance goal (P < 0.0001). Procedural success was 97.0% (324 of 334; exact 2-sided 95% CI: 94.6%-98.6%); the lower 1-sided 97.5% confidence bound was 94.6%, above the 87.1% performance goal (P < 0.0001). In the OCT substudy, mean calcium length was 22.2 ± 10.3 mm, and the maximum contiguous calcium site had a calcium arc of 247.3 ± 99.0° and a thickness of 0.90 ± 0.25 mm at baseline. At this site, final stent area was 7.1 ± 1.9 mm2, with stent expansion of 95.9% ± 18.6%. Calcium fracture was identified in 62.5% (50 of 80) of lesions after intravascular lithotripsy and 66.3% (53 of 80) after stenting.

CONCLUSIONS: TECTONIC CAD met both prespecified coprimary safety and effectiveness endpoints, demonstrating low 30-day target lesion failure and high procedural success. OCT demonstrated calcium fracture and favorable stent expansion at maximum contiguous calcium.

Last updated on 11/03/2026
PubMed